MCS100FT Multi gas analyser | EMS
MCS100FT Multi gas analyser

Extractive FTIR based multi-gas analyser

The FTIR measuring principle allows the simultaneous determination of more than 12 measuring components – tailor-made for the particular requirements of the customer. The MCS100FT FTIR analyser system with its heated measuring cell enables the monitoring of hydrogen fluoride limits as requested by legislation. The MCS100FT is equipped with an oxygen sensor as standard and can be supplemented with a total hydrocarbon analyser. With its reliable measuring technology, its easy operation and low maintenance requirements, the MCS100FT provides a solution that offers unequalled opportunities.
Measured valuesCH4, CO, CO2, Corg, HCl, HF, H2O, NH3, NO, NO2, N2O, O2, SO2, NOx, C3H8, C2H6
Performance-tested measurementsCH4, CO, CO2, Corg, HCl, HF, H2O, NH3, NO, NO2, N2O, O2, SO2
Measurement principleFTIR spectroscopy, flame ionisation detection, Zirconium dioxide sensor
Gas flow rate ≤ 300 l/h
Measuring rangesCH4 0 ... 70 ppm / 0 ... 210 ppm CO 0 ... 60 ppm / 0 ... 1,200 ppm CO2 0 ... 25 Vol.-% Corg 0 ... 7.5 ppm / 0 ... 75 ppm HCl 0 ... 10 ppm / 0 ... 100 ppm HF 0 ... 3 ppm / 0 ... 10 ppm H2O 0 ... 40 Vol.-% NH3 0 ... 13 ppm / 0 ... 65 ppm NO 0 ... 150 ppm / 0 ... 1,500 ppm NO2 0 ... 25 ppm / 0 ... 250 ppm N2O 0 ... 25 ppm / 0 ... 250 ppm O2 0 ... 21 Vol.-% SO2 0 ... 25 ppm / 0 ... 525 ppm NOx 0 ... 100 ppm / 0 ... 1,000 ppm C3H8 0 ... 25 ppm C2H6 0 ... 40 ppm Other measuring ranges and components on request Certified measuring ranges CH4 0 ... 50 mg/m³ / 0 ... 150 mg/m³ CO 0 ... 75 mg/m³ / 0 ... 300 mg/m³ / 0 ... 1,500 mg/m³ CO2 0 ... 25 Vol.-% Corg 0 ... 15 mg/m³ / 0 ... 50 mg/m³ / 0 ... 150 mg/m³ / 0 ... 500 mg/m³ HCl 0 ... 15 mg/m³ / 0 ... 90 mg/m³ / 0 ... 150 mg/m³ HF 0 ... 3 mg/m³ / 0 ... 10 mg/m³ H2O 0 ... 40 Vol.-% NH3 0 ... 10 mg/m³ / 0 ... 50 mg/m³ NO 0 ... 200 mg/m³ / 0 ... 400 mg/m³ / 0 ... 2,000 mg/m³ NO2 0 ... 100 mg/m³ / 0 ... 500 mg/m³ N2O 0 ... 50 mg/m³ / 0 ... 500 mg/m³ O2 0 ... 21 Vol.-% SO2 0 ... 75 mg/m³ / 0 ... 300 mg/m³ / 0 ... 1,500 mg/m³
Response time (t90)≤ 200 s FID ≤ 45 s
Sensitivity drift< 3 %: of the measuring range full scale value per maintenance interval FID < 2 %: of measuring range full scale per week
Zero point drift < 3 %: of the measuring range full scale value per maintenance interval FID < 2 %: of measuring range full scale per week
Detection limit< 2 %: of measuring range full scale
Process temperature≤ +1,300 °C Sample gas temperature: ≤ +220 °C
Process pressure900 hPa ... 1,100 hPa
Ambient temperature+5 °C ... +35 °C, With cooling device +5 °C ... +50 °C
Storage temperature–20 °C ... +60 °C
Ambient pressure900 hPa ... 1,100 hPa
Ambient humidity ≤ 80 % Non-condensing
Conformities-Approved for plants requiring approval2001/80/EC (13. BImSchV) 2000/76/EC (17. BImSchV) 27.BImSchV TA-Luft (Prevention of Air Pollution) EN 15267 EN 14181 MCERTS
Electrical safetyCE
Enclosure ratingIP43, Optional IP54